JM Modern HikaShop - шаблон joomla Видео
(06-1) 619-8321
Nuaire

Közvetlen fűtésű önsterilizáló 145˚C, CO2/RH, NU-5820

nu-5820-interior-1
nu-5820-interior-2
Az In-VitroCell ES NU-5820 Mikrobiológiai CO2 inkubátor jellegzetessége a relatív páratartalom (RH) szabályozás képessége, mely legfőképpen a kiszáradásra hajlamos sejtvonalak esetében előnyös tulajdonság. A készülékbe szerelt érzékelő folyamatosan méri a kamra páratartalmát és a beállított érték fenntartása érdeképen 99,99% hatékonyságú HEPA filter által szűrt gőzt juttat a kamrába.

ÁRAJÁNLATOT, INFORMÁCIÓT KÉREK!

Relative Humidity (RH) Control

A RH sensor continuously samples chamber humidity levels to monitor and control the humidity set point. If the system senses any variation in humidity, vapor is injected into the chamber by a water reservoir located underneath the chamber. 

 

Dual Sterilization Cycles

Easily accessible through the NuTouch Intelligent Interface, In-VitroCell offers 145°C high heat sterilization and a 95°C humidified decontamination cycles as a preventative measures and to eradicate contamination in the growth chamber. On screen instructions read users through the process. A timer function is available to delay the sterilization process for an overnight procedure. 

 

NuTouch Intelligent Interface

NuTouch is a user-friendly 5” x 7” (127 x 178 mm) color touch screen designed for high user adoption. The electronic control system is a microcomputer designed to service the precise control requirement of the chamber environment for culture growth. The microcomputer is supported with Read Only Memory (ROM) containing executable software, Random Access Memory (RAM) for temporary storage, and Electronically Erasable Programmable Read Only Memory (EEPROM) for control set points and parameters. The EEPROM provides for indefinite storage of these values during periods of power off or power interruption (power fault tolerant). NuTouch features are:

  • Temperature and Gas Monitor and Control
  • Status Indicators – Screen Identity; Run or Standby mode; Door Adjar; Air and CO2 Inject; Air and CO2 Chamber Sampling
  • Date and Time Functions in International Format
  • Password Protection
  • Onscreen Icon Descriptions – Pressing on any icon will provide a description of what the item is.
  • Onscreen Directions – Directions are available to perform calibrations.
  • Step-by-Step directions - Onscreen directions asisit with calibrations and sterilization cycles.
  • Alarm Status Menu - A red “Alarm Status” text button will appear on the Main Screen to indicate a fault within the Incubator system. 
  • Maintenance Required Reminders – Onscreen Icon “MAINTENACE REQUIRED” appears when a programmable reminder (0 to 48 months) hits set point to notify filter check or when the RH water resviour needs to be refilled.
  • Historical Performance Monitor – Temperature and CO2 gas history is available to view onscreen in graphical form or can be downloaded via USB for alarms, control systems, events, and performance information.
  • Service Settings and Complete Diagnostic Controls
  • System Information
  • Display Settings
  • Available in English, French, German, and Spanish

Constant Contamination Control (C3)

A unique blend of product features to promote cell growth while minimizing potential contamination creating the most intrinsic and reliable growth chamber.

Closed Loop HEPA Filtration

A maintenance free air pump continuously draws environmental samples from two locations inside the chamber and passes the sample through a 99.99% HEPA filter before entering the sensor bay to measure gas and humidity levels. The IR sensor measures CO2 gas and humidity levels from the sample to monitor before injecting back into the growth chamber. The NuTouch Electronic Control system will then determine the appropriate course of action to maintain chamber set-points by placing the system on standby or injecting 99.99% HEPA filtered CO2 gas, air, or vapor into the growth environment.

ISO Class 5 Cleanroom Conditions

Based on clean room technology, In-VitroCell maintains the growth chamber at positive pressure similar to an ISO Class 5 Clean room to slow airflow to minimize desiccation. Anytime the chamber door is open to the laboratory environment, clean HEPA filtered will be forced out instead of contaminated lab air being drawn in. 

 

Single Beam, Dual Wave Infrared (IR) Sensor

A microprocessor-based, non-dispersive, single source dual wave infrared (IR) sensor monitors and controls CO2 levels inside the growth chamber. Wavelengths are only absorbed by CO2 making measurement insensitive to other components, such as water vapor. The stable IR sensor has a controlled range of 0.1 to 20% accurate within 0.1%. 

Temperature Uniformity

The chamber walls are surrounded by five (5) heating elements to control temperature. A door heater is separately controlled to minimize condensation on the inner chamber door. The heating elements are wrapped in a high density R5 insulation to maintain temperature integrity. Dual temperature sensor probes monitor and control temperature set points to ensure consistency and uniformity throughout the chamber. 

 

Advanced Construction

The inner chamber is constructed of high grade 16 gauge, type 304L polished stainless steel using crevice-free construction that provides an easily cleanable inert surface that does NOT promote biological growth. Seamless coved interior corners are easily cleaned promoting a healthy growth chamber. A single piece gasket creates a liquid tight seal from incubator chamber to inner door to eliminate condensation build up and potential contamination. Inner chamber shelving and gasket may be removed and routinely autoclaved if desired. 

 

Minimal Vibration Promotes Cell Growth

By minimizes moving parts, In-VitroCell ES utilizes an accurate air pump to inject air and gas to eliminate potential vibration.

 

Smaller Footprint with More Useable Space

A large 7 cubic foot (200 Litre) growth chamber comes standard with four incubator shelves with a maximum capacity of 23 shelves. By attaching heating elements directly to the chamber and insulating with High Density R5 insulation In-VitroCell was engineered as a compact design while maximizing chamber capacity.

 

CuVerro® Antimicrobial Copper Surfaces (Optional Feature)

Add CuVerro® Antimicrobial Copper Surfaces to the incubator growth chamber and/or shelving to kill bacteria* to minimize potential incubator contamination. CuVerro® is laboratory tested and EPA registered. CuVerro® Antimicrobial Copper Surfaces kill more than 99.9% of bacteria* within 2 hours, and continues to kill 99% of bacteria* even after repeated contamination, when cleaned regularly.

 

EPA Reg No 85353-5

 

EPA Est No 088257-MN-001

 

*Laboratory testing shows that, when cleaned regularly, CuVerro® antimicrobial copper surfaces kill greater than 99.9% of the following bacteria within 2 hours of exposure: MRSA, Staphylococcus aureus, Enterobacter aerogenes, Pseudomonas aeruginosa, and E. coli O157:H7. CuVerro® antimicrobial copper surfaces are a supplement to and not a substitute for standard infection control practices and have been shown to reduce microbial contamination, but do not necessarily prevent cross contamination; users must continue to follow all current infection control practices, including those practices related to cleaning and disinfection of environmental surfaces.

 

Standard features

Four (4) Stainless Steel Shelves

Eight (8) Stainless Steel

Shelf Brackets

Shelf Guides

Right Hinge Door Swing

Remote Alarm Output Contacts

4 to 20 mA Analog Output

RS-485 Communication

USB Port

CO2 Sample Port

Adjustable Leg Levelers

Access Port and plug with breather holes

One (1) Water Pan

One (1) 6.5 ft. / 2 m Electrical Cord

Műszaki specifikációk

Model

NU-5820

Chamber Volume 7.06 ft.3 / 200 L
Exterior Dimensions (W x D x H) 
in. [mm]
26.75 x 27.0 x 41.5 [679.5 x 691 x 1054
Chamber Dimensions (W x D x H) 
in. [mm]
21.15 x 21 x 26.5 [546.1 x 533.4 x 673.1]
Weight 235 lbs. / 106 kg.

Shelf Capacity

Dimensions (W x D) in. [mm] 19 ¾ x 18 ¾ [502 x 476]
Supplied 4 Shelves
Max. Capacity 23 Shelves
Max. Weight Capacity 20 lbs. (9 kg) per Shelf
  50 lbs. (23 kg) per Incubator

Utility Connections

Gas Connections 0.25" (6.3 mm) Tubing
Gas Input Pressure 20 PSIG (1.4 BAR) Input Pressure Maximum. Two Stage Gas Regulator Required

Electrical Requirements

Startup Power 625 watts
Running Power 250 watts
Sterile/Decon Cycle 995 watts
Heat Rejected 14 BTU / minute

Temperature Control System

Sensor Type Precision Integrated Circuit
Range 5°C to 55°C (5°C aboveambient to 30°C max. ambient)
Accuracy ±0.1°C
Uniformity ± 0.3°C at 37°C
Recovery 0.12°C/minute Average
Display Resolution 0.1°C
 Minimum Qualifications for Sterilization
145 DEG Cycle 135°C
95 DEG Cycle 85°C

CO2 Control System

Sensor Type Infrared single source dual wave length
Range 0.1 to 20%. (0.0 set point idles system)
Accuracy ± 0.1%
Recovery Up to 5% ±0.2% / -0.5 in 5 minutes average
Display Resolution ± 0.1%
CO2 Control Logic Fixed/Algorithm/Manual, Environmental Adaptable

Relative Humidity (RH) Control System

Default Set-point 90%
Range 5% above ambient to 90%
Accuracy ± 3%
Recovery 90% ±3% / 25 minutes

Quality Asssurance

ULUL-C, CE, Energy Saver, RoHS, Constant Contamination Control, Critical Cleaning Components
Applikációk
    • RS-485 Communication for In-VitroCell Incubators
    • 4-20mA Analog Output for In-VitroCell CO2 Incubators
    • Improve the Performance of your CO2 Incubator
    • CuVerro® Care and Use

      NuAire products manufactured with CuVerro® have specific care and use instructions to assure maximum customer satisfaction. CuVerro® is a class of copper based alloys that continuously kill bacteria associated with life science laboratory contamination. These surfaces represent the only class of EPA-registered solid surface materials that actively kill bacteria. CuVerro® is suited for common surfaces where contamination is a concern. Laboratory surfaces benefitting from CuVerro® are inherently any work surface that can harbor contamination. NuAire fabricates CuVerro® in many ways to produce a variety of products for the laboratory. Understanding the care and use of CuVerro® will maximize its benefit and visual appearance. Find additional information at www.cuverro.com.

      CuVerro® may be cleaned with various laboratory surface disinfectants by different methods. For all surface disinfectants, follow application and use instructions provided by the manufacturer. Download the whitepaper to view a list of surface disinfectant types and their visual effect on CuVerro®. --- CuVerro® Care and Use

Tartozékok, kiszerelések
  • IsoBar Surge Suppressor

  • CO2 Regulator Two Stage

    CO2 Regulator Two Stage

  • Dry In-Line Filter

    Dry In-Line Filter

  • Wet In-Line Filter

    Wet In-Line Filter

  • Refill CO2 Analyzer Fluid Kit

    Refill CO2 Analyzer Fluid Kit

  • CO2 Analyzer Fyrite Kit

    CO2 Analyzer Fyrite Kit

  • External Tank Switch

    External Tank Switch

  • CO2 Regulator, Single Stage

    CO2 Regulator, Single Stage

  • CO2 Incubator Tubing

    CO2 Incubator Tubing

  • Mobile Platform with Leg Leveling Casters and Mounting Brackets for In-VitroCell ES CO2 Incubators

    Mobile Platform with Leg Leveling Casters and Mounting Brackets for In-VitroCell ES CO2 Incubators

  • CO2 Incubator Tank Alarm

    CO2 Incubator Tank Alarm

  • Stainless Steel Incubator Shelf and Guide Brackets (In-VitroCell ES Series)

    Stainless Steel Incubator Shelf and Guide Brackets (In-VitroCell ES Series)

  • Incubator Water Pan (In-VitroCell ES Series)

    Incubator Water Pan (In-VitroCell ES Series)

  • CuVerro® Antimicrobial Copper CO2 Incubator Shelf and Guide Brackets (In-VitroCell ES Series)

    CuVerro® Antimicrobial Copper CO2 Incubator Shelf and Guide Brackets (In-VitroCell ES Series)

blue down arrowForgalmazott termékeink gyártói