Safety is Priority Number 1 |
||
![]() |
||
True Laminar Airflow |
||
|
Unidirectional airflow moving along parallel flow lines at a constant velocity of 60 fpm (0.30 m/s) minimizes air turbulence within the work zone lessening the chance for cross contamination. |
||
![]() |
||
Dynamic Air Barrier |
||
|
A strong inflow air velocity of 105 fpm (0.53 m/s) over an aerodynamic airfoil creates a dynamic air barrier offering maximum personnel protection. |
Ergonomic Design Reduces Workplace Injuries |
||
|
LabGard® offers a shorter reach into the work zone bringing work materials closer to you relieving stress on your arms, back and neck. |
||
|
A standard plastic armrest eliminates cold conduction on forearms relieving tendons of any unnecessary stress. |
![]() |
|
|
A recessed airfoil allows technicians to use the armrest without fear of blocking airflow. |
||
|
A further recessed work-tray with spill lip protection lowers the work zone making it easier for lab techs to use pipettes or creating additional space for machinery such as a liquid handlers or TPN Compounding Systems. |
||
|
A large user friendly work zone offers more usable space while making the farthest corners easier to reach. |
||
Long Lasting NuAire® Construction |
||
No Two Cabinets Are Created Equal |
||
|
LabGard® Biological Safety Cabinets are constructed from 16/18 gauge, type 304 stainless steel welded together as a uniform monolithic shell. The work zone features a coved back-wall insert welded into place creating double wall construction adding to the cabinet's integrity. LabGard® is Silicone Free and designed to provide long lasting safety. |
Technology |
||
![]() |
||
AeroMax™ Electronic Control System |
||
|
AeroMax™ ECS provides a user friendly control systems that monitors and displays airflow performance via PressureFlow™ - A Digital Pressure Monitor. nitecare™ Night Setback Mode - Upon sliding the window sash into the closed position, the motor blower will continue to operate at a lower rate to save energy and maintain interior clean air conditions ready for use upon window sash opening. |
||
Ultra High Efficiency Motor |
||
|
An ultra-high efficiency ECM motor combined with the properly sized fan blade lowers energy usage while extending the life of filters. LabGard® offers the lowest total cost of ownership compared to other models. |
![]() |
|
![]() |
||
HEPEX™ Zero Leak Airflow System |
||
|
All LabGard® positive pressure plenums are surrounded by a vacuum relative to the room. Meaning in the event of a leak via gasket failure, all contaminated air in the plenum is still forced through HEPA filtration. The HEPEX™ system also works to provide true laminar airflow by evenly distributing plenum air over the supply HEPA filter. The flexible design eliminates any vibration transfer from the motor blower as well as creates a serviceable work environment for Certifiers after decontamination. |
||
Make it Easier on your Service Team |
||
|
We at NuAire® have designed each LabGard®Biosafety Cabinet to be service friendly. 95% of service functions can be performed at the front of the cabinet. A flip down control panel creates a workable tray for service technicians. Access filters and the motor blower from the front of the cabinet with the removal of a safety panel. |
![]() |
|
|
When it comes time to any type of internal service make it easy with our DECON101™ System. TheDECON101™ System works with a variety of decontamination systems using Formaldehyde, Chlorine Dioxide or Vapor Phase Hydrogen Peroxide. |
![]() |
|
The NuAire® Advantage |
||
|
NuAire® utilizes an independent network of sales and service professionals offering our customers a vast array of knowledge and relationships. Our inside sales and service teams are the pride of the industry; some of which have been with NuAire® since the beginning in 1971. We work hard and diligently to make you happy. That is our mission and that is why we come into work everyday. Purchase a NuAire® and become part of our family. |
NU-540-300 |
NU-540-400 |
NU-540-500 |
NU-540-600 | |
|---|---|---|---|---|
Nominal Width |
3 Foot (0.9 m)
|
4 Foot (1.2 m)
|
5 Foot (1.5 m)
|
6 Foot (1.8 m)
|
Performance Specifications
|
NSF/ANSI 49 EN 12469 |
|||
NSF/ANSI 49 |
Class II
|
|||
Style of Cabinet |
Bench Top or Console with Base Stand
|
|||
Construction |
All welded stainless steel 16/18 Gauge, Type 304 pressure tight design
|
|||
Diffuser for Air Supply (Metal) |
Non-fammable
|
|||
HEPA Filter Seal Type: |
||||
|
Supply Filter 99.99% @ 0.3 Micron Exhaust Filter 99.99% @ 0.3 Micron |
HEPEX Seal Neoprene, Spring Loaded
|
|||
Fumigation per NIH/NSF Procedures |
Yes
|
|||
Standard Services |
||||
| Service Coupling (3/8" [9.525 mm] NPT) |
One Right Sidewall Two Left Sidewall |
|||
| Service Valve (3/8" [9.525 mm] NPT) |
One Right Sidewall
|
|||
| Outlet(s) |
Two Backwall
|
|||
Optional Services |
||||
| Service Valve (3/8" [9.525 mm] NPT) |
Up to Three each Sidewall
|
|||
| Ultraviolet Light |
One Backwall
|
|||
| Standard / Cup Sinks |
Left or Right Worksurface
|
|||
Work Access Opening |
||||
| Standard Opening Height |
10" (254 mm)
|
|||
| Optional Opening Height |
8" (203 mm)
|
|||
| Standard Inflow Velocity |
105 fpm (0.53 m/s)
|
|||
Viewing Window |
Standard Safety Plate Sliding Glass Fully Closed to 21" (533 mm) Open
|
|||
Exterior Dims. |
||||
| Height | 61.875" (1572 mm) | 61.875" (1572 mm) | 61.875" (1572 mm) | 61.875" (1572 mm) |
| Height (min) | 59" (1499 mm) | 59" (1499 mm) | 59" (1499 mm) | 59" (1499 mm) |
| Width | 41.625" (1057 mm) | 53.625" (1362 mm) | 65.625" (1669 mm) | 77.625" (1972 mm) |
| Depth (Armrest Removed) | 31.4375" (799 mm) | 31.4375" (799 mm) | 31.4375" (799 mm) | 31.4375" (799 mm) |
Interior Dims. |
||||
| Height | 28.5" (724 mm) | 28.5" (724 mm) | 28.5" (724 mm) | 28.5" (724 mm) |
| Width | 34.375" (873 mm) | 46.375" (1178 mm) | 58.375" (1483 mm) | 70.375" (1788 mm) |
| Depth***** | 25.75" (654) | 25.75" (654) | 25.75" (654) | 25.75" (654) |
Required Exhaust |
||||
10" (254 mm) access opening |
||||
| Canopy Variable Flow Thimble (NU-911) |
276 - 501 CFM (649 - 851 CMH) |
363 - 588 CFM (617 - 1000 CMH) |
451 - 676 CFM (766 - 1149 CMH) |
538 - 763 CFM (915 - 1297 CMH) |
| Canopy Fixed Flow Thimble (NU-907) | 320 CFM (544 CMH) | 426 CFM (724 CMH) | 531 CFM (445 CMH) | 634 CFM (1077 CMH) |
8" (203 mm) Access Opening |
||||
| Canopy Variable Flow Thimble (NU-911) |
255 - 450 CFM (383 - 765 CMH) |
295 - 520 CFM (502 - 884 CMH) |
365 - 590 CFM (621 - 1003 CMH) |
436 - 661 CFM (741 - 1124CMH) |
| Canopy Fixed Flow Thimble (NU-907) | 269 CFM (457 CMH) | 359 CFM (610 CMH) | 445 CFM (756 CMH) | 532 CFM (904 CMH) |
Plant Duct Static Pressure |
0.05 ‐ 0.1" / 1.27 ‐ 2.54mm H2O | |||
Heat Rejected, BTU, Per Hour |
||||
Non-Vented |
||||
| 10" (254 mm) access opening | 903 | 1140 | 1768 | 1884 |
| 8" (203 mm) access opening | 826 | 1020 | 1611 | 1768 |
Vented |
120 | 157 | 198 | 198 |
Electrical |
||||
115 V, 60 Hz |
||||
| Amps | ||||
| Blower/Lights (10 / 8 Access opening)** | 2.3 / 2.1 | 2.9 / 2.6 | 4.5 / 4.1 | 4.8 / 4.5 |
| Duplex | 3 | 3 | 3 | 3 |
| Rated | 10 | 12 | 14 / 12* | 14 / 12* |
| 12 ft. Power Cord (one) | 14 GA - 3 Wire, 15A | 14 GA - 3 Wire, 15A |
14 GA - 3 Wire, 20A Optional: 12 GA - 3 Wire 15 A* |
14 GA - 3 Wire, 20A Optional: 12 GA - 3 Wire 15 A* |
230 V, 50/60 Hz |
||||
| Amps | ||||
| Blower/Lights (10 Opening)** | 2.3 | 2.9 | 4.5 | 4.8 |
| Duplex | 3 | 3 | 3 | 3 |
| Rated | 8 | 10 | 11 | 11 |
| 12 ft. Power Cord (one) | 14 GA - 3 Wire, 15A | 14 GA - 3 Wire, 15A | 14 GA - 3 Wire, 15A | 14 GA - 3 Wire, 15A |
Weight |
||||
| Crated Shipping Weight*** | 450 lbs. / 204 kg. | 515 lbs. / 234 kg. | 600 lbs. / 272 kg. | 670 lbs. / 304 kg. |
| Net Weight | 400 lbs. / 181 kg. | 465 lbs. / 211 kg. | 550 lbs. / 249 kg. | 620 lbs. / 281 kg. |
Sound Pressure Level per ISO 4871**** |
Not to exceed 55 dbA | Not to exceed 56 dbA | Not to exceed 58 dbA | Not to exceed 60 dbA |
|
*15A Configuration does not include accessory outlet. **Based on cabinet with new filters ***Crated shipping weight does not include weight for accessories or options ****Uncertainty is K = 2 dbA, measurement performed per ISO 11201 in normal running mode. ***** Measured at 10" (254 mm) Window Height |
||||
Elbow Rest (10")
Elbow Rest (5")
Ergonomic Accessories Package
Stainless Steel Turntable
TechniSat 9" x 11" (22.9cm x 27.9cm) pre-wetted with 70% IPA Wipes
SterileWipe LP 9" x 9" (23 cm x 23 cm) Sterile Wipes
Sterile AlphaSat® with AlphaSorb® HC 12" x 12" (31 cm x 31 cm) 125 Wipes
Sterile AlphaSat® 10 folded polyester wipers pre-wetted with 70% IPA (ISO Class 3 - 5)
TechniCloth® Wipers
Mini AlphaMop™ AlphaWipe® Polyester Replacement Mop Covers
Foam Armrest Pads - Package Of 10
Alnor AirGard BSC Airflow Monitor, 115V/60Hz
Alnor AirGard International BSC Airflow Monitor, 230V/50Hz
AireValv Manual Airtight Butterfly Valve 8" Diameter
AireValv Manual Airtight Butterfly Valve 10" Diameter
AireValv Manual Airtight Butterfly Valve 12" Diameter
Silicone Connection Sleeve Kit for all Models
Round to Round Transition, Duct to Hood transition connection - 8" to 10"
Flex Duct Kit 6" Diameter
AireValv Manual Butterfly Damper 6" Diameter
Alnor AirGard High and Low BSC Airflow Monitor
Digital Manometer Monitor 115V
High/Low LED Bar Pressure Monitor 115V/60Hz
Digital Manometer Monitor 230V
High/Low LED Bar Pressure Monitor 230V
Flex Duct Kit 8" Diameter
Flex Duct Kit 10" Diameter
Flex Duct Kit 12" Diameter
Disposable Foam Armrest Pads - Package of 3
Forgalmazott termékeink gyártói - keressen gyártó szerint a logóra kattintva