The densitometers are designed and factory calibrated to measure turbidity in the range of 0.5 to 4.0 McFarland units (DEN-1) 0.0 to >7.5 McFarland units (DEN-1B) with a small standard deviation. If required, it can deliver a wider
measurement range (up to 15.00 McFarland units), but with a greater standard deviation.
| Densitometers - specification | |||
|---|---|---|---|
| Densitometer | |||
| DEN-1 | DEN-1B | ||
| Light source | light diode | ||
| Wavelength | nm | 565 ± 15 | |
| Range | McFarland units | 0.3 to 15.00 | 0.00 to 15.00 |
| Standard deviation, McFarland unit | 0.0 McF | 0.0 ± 0.1 | |
| 0.5 McF | 0.5 ± 0.1 | ||
| 3.0 McF | 3.0 ± 0.1 | ||
| 6.0 McF | 6.0 ± 0.2 | ||
| ≥ 7.5 McF | 7.5 ± 0.2 | ||
| Precision | ± 3% | ||
| Measurement time | sec | 1 | |
| Tube diameter | external diameter, mm | 18 | |
| with D-16 adaptor , mm | 16 | ||
| Sample volume | ml | ≥ 2 | |
| Display | 2 digit LED | LCD / 0.01 McF | |
| External power supply | 12V, 300 mA | ||
| Independant power supply | - | 3 x AAA batteries | |
| Ambient temperature range | °C | 4 to 40 | |
| Dimensions, w x d x h | mm | 165 x 115 x 75 | |
| Weight | kg | 0.9 | |
Life science - typical applications include determining concentration of cells (bacterial and yeast cells) in the fermentation process, detecting the susceptibility of micro-organisms to antibiotics, identifying micro-organisms with various test systems, and measuring optical density at fixed wavelength.
Options and accessories | |
|---|---|
| D-16 | tube adaptor for tubes with 16 mm outer diameter. |
| DEN MCF STDS | set of liquid McFarland standards in 16mm ø glass tubes. 0.5 / 1.0 / 2.0 / 3.0 / 4.0 can be decanted into an alternative tube (shelf-life no longer valid). Requires D16 adaptor, supplied with DEN-1 / DEN-1B. |
| DEN MCF 18 STDS | set of gel McFarland standards in 18mm ø glass tubes. 0.0 / 0/5 / 3.0 / 6.0 / >7.5. |